Multi-use Temperature and Humidity Logger
RC-51H User Manual Multi-use Temperature&Humidity Data Logger
Product Overview
This temperature and humidity data logger is mainly used in the fields or places of medicine, food, life science, flowers breeding industry, ice chest, container, shady cabinet, medical cabinet, refrigerator, laboratory, and greenhouse, etc. RC-51H is plug-and-play and it can directly generate the data report, with no need to install data management software. The data can still be read in case battery runs out.
Structure Description
1 | Transparent cap | 5 | Button & Bi-color indicator (red and green) |
2 | USB port | ||
3 | LCD screen | 6 | Sensor |
4 | Seal ring | 7 | Product label |
LCD screen
A | Battery indicator | H | Humidity unit or Progress percentage |
B | Mean kinetic temperature | ||
C | Start recording indicator | I | Timing indicator |
D | Stop recording indicator | J | Average value indicator |
E | Cyclic recording indicator | K | Number of records |
F | Computer connection indicator | L | Combined indicator |
G | Temperature unit(°C/°F) |
For more details, please refer to the menu and status indicator
Product label(I)
a | Model | d | Barcode |
b | Firmware version | e | Serial number |
c | Certification information |
: Picture is for reference only, please take real object as standard.
Technical Specifications
Recording Options | Multi-Use |
Temperature Range | -30°C to 70°C |
humidity Range | 10%~95% |
Temperature & humidity Accuracy | ±0.5(-20°C/+40°C);±1.0(other range) ±3%RH (25°C, 20%~90%RH), ±5%RH(other range) |
Data Storage Capacity | 32,000 readings |
Software | PDF/ElitechLog Win or Mac (latest version) |
Connection Interface | USB 2.0, A-Type |
Shelf Life/Battery | 2 years1/ER14250 button cell |
Recording Interval | 15 minutes(standard) |
Startup Mode | Button or software |
Stop Mode | Button, software or stop when full |
Weight | 60g |
Certifications | EN12830, CE, RoHS |
Validation Certificate | Hardcopy |
Report Generation | Automatic PDF report |
Temperature & humidity Resolution | 0.1°C(Temperature) 0.1%RH (Humidity) |
Password Protection | Optional on request |
Reprogrammable | With free Elitech Win or MAC software |
Alarm Configuration | Optional, up to 5 points, Humidity only supports upper and lower limit alarm |
Dimensions | 131 mmx24mmx7mm(LxD) |
1.Depending on optimal storage conditions (±15°C to +23°C/45% to 75% rH) |
Software download: www.elitecilus.com/download/software
Parameter Instruction
Users can reconfigure the parameters by data management software per actual needs. The the original parameters and ata will be cleared.
Alarm threshold | This data logger supports 3 upper temperature limits, 2 lower temperature limits, 1 upper humidity limit and 1 lower humidity limit. | |
Alarm zone | The zone which beyond the alarm threshold | |
Alarm type | Single | The data logger records the single time for continuous over-temperature events. |
Cumulative | The data logger records the cumulative time of all the over-temperature events. | |
Alarm delay | The data logger does not alarm immediately when the temperature is within the alarm zone. It begins to alarm only when the over-temperature time elapses the alarm delay time. | |
MKT | Mean kinetic temperature, which is an evaluation method of the temperature fluctuation effect on the goods in storage. |
Operating Instructions
This data logger can be stopped by software. Users can stop the logger by clicking the stop button in the data management software.
Action | Parameter configuration | Operation | LCD indicator | Indicator |
Start | Instant-on | Disconnect to USB | Green indicator flashes 5 times. | |
Timing start | Disconnect to USB | Green indicator flashes 5 times. | ||
Manual start | Press and hold for 5s | Green indicator flashes 5 times. | ||
Manual start (delayed ) | Press and hold for 5s | Green indicator flashes 5 times. | ||
Stop | Manual stop | Press and hold for 5s | Red indicator flashes 5 times. | |
Over-Max-record-capacity stop (disable manual stop) | Reach the Max capacity | Red indicator flashes 5 times. | ||
Over-Max-record-capacity stop (Enable manual stop) | Reach the Max capacity or press and hold the button for 5s | Red indicator flashes 5 times. | ||
View | Press and release the button Refer to the menu and status indicator |
View data When the data logger is inserted into the USB port of the computer, the data report will be created automatically. The red and green indicators flash in turn when the document is being created, and the LCD screen shows the progress of PDF Report creation. The red and green indicators light at the same time immediately after the document is created, then users can view the data report. The document creation will last for no more than 4 minutes.
(1) Rotate the transparent cap in the direction of the arrow and remove it.
(2) Insert the data logger into the computer and view the data report.
Software download: www.elitechus.com/download/software
Menu and Status Indicator
Description of the indicator flashing status
Status | Action of the indicators |
Not started | The red and green indicators flash 2 times simultaneously. |
Start delay Timing | The red and green indicators flash once simultaneously. |
Started-normal | The green indicator flashes once. T |
Started-alarm | The red indicator flashes once. T |
Stopped-normal | The green light flashes 2 times. |
Stopped-alarm | The red light flashes 2 times. |
Description of the menus
Menu | Description | Example |
11 | Countdown of the (timing) start | |
Countdown of the (delayed) start | ||
2 | Current temperature value | |
3 | Current humidity value | |
4 | Points of the records | |
5 | Average temperature value | |
6 | Average humidity value | |
7 | Maximum temperature value | |
8 | Maximum humidity value | |
9 | Minimum temperature value | |
10 | Minimum humidity value |
Description of the combined indicators and other status
Display | Description |
(group)³ | No alarm |
(group) | Already alarmed |
(group) | Minimum value |
(group) | Maximum value |
(group) rotating | Rate of progress |
Null value | |
Clear data | |
In USB communication |
Note: 1 Menu 1 appears only when the corresponding function is selected.
2 “” should be in a state of blinking.
3 The display in the combined indicator area. The same as below.
Replace battery
(1) Press the bayonet in the direction of the arrow and remove battery cover
(2) Place a new battery
(3) Install the battery cover in the direction of the arrow
Software download: www.elitechus.com/download/software
Report
The first page Other pages
1 | Basic information |
2 | Description of the usage |
3 | Configuration information |
4 | Alarm threshold and related statistics |
5 | Statistical information |
6 | Temperature and humidity graph |
7 | Temperature and humidity data details |
A | Document creation time (record stop time) |
B | Alarm (Alarm status as shown in the figure above) |
C | Stop mode that has been set. |
D | Alarm status of the temperature alarm zone |
E | Total times of exceeding the temperature alarm threshold |
F | Total time of exceeding the temperature alarm threshold |
G | Alarm delay and alarm type |
H | Alarm threshold and temperature alarm zones |
I | Actual stop mode (different from the item C) |
J | Vertical coordinate unit of the data graph |
K | Alarm threshold line (corresponding to the item L) |
L | Alarm threshold |
M | Record data curve (black indicates temperature, deep green indicates humidity) |
N | Document name (serial number & description of usage ID) |
O | Record time range in the current page |
P | Records when date changes (date & temperature and humidity) |
Q | Records when the date is not changed (time & temperature and humidity) |
Attention: The data above is only used as explanation of the report. Please refer to the actual document for specific configuration and information.
What’s included
1 temperature and humidity data logger | 1 Er14250 battery | 1 user manual |
Elitech Technology, Inc.
www.elitechus.com
1551 McCarthy Blvd, Suite 112
Milpitas, CA 95035 USA V2.0
Software download: www.elitechus.com/download/software